Code: 000555
Spectrophotometer, UV-VIS, single beam
Wavelength Range: 190-1100nm
Spectral Bandwidth: 2nm(5nm, 1nm optional)
Wavelength Accuracy: ±0.5nm
Wavelength Reproducibility: 0.2nm
Photometric Accuracy: ±0.5%T(0-100%T), ±0.002A(0-0.5A), ±0.004A(0.5A-1A)
Photometric Reproducibility: 0.2%T
Working Mode: T, A(-0.33A), C, E
Stray Light: ≤0.1%T(NaI, 220nm; NaNO2, 340nm)
Baseline Flatness: ±0.002A
Stability: ≤0.002A/h( at 500nm, after warming up)
Noise: ±0.001A (at 500nm, after warming up)
Detector: Silicon photo-diode
Display: 6 inches high light blue LCD
Power: AC:220V/50Hz, 110V/60Hz, 140W
Dimensions: 530×410×210mm
Weight: 18kg
On Request