The site uses cookies. By clicking on “Accept”, automatically you approve the use of cookies as stated at terms of use.
ACCEPT COOKIES

Code: 001320

X, γ Radiation personal dosimeter, portable




<br />


























































Individual dose equivalent measurement range

 

AT3509, AT3509A Hp(10)

1 µSv...10 Sv

AT3509B Hp(10), Hp(0.07)

1 µSv...10 Sv

AT3509C Hp(10), Hp(0.07)

1 µSv...10 Sv

Individual dose rate equivalent measurement range

 

AT3509, AT3509A Ḣp(10)

0.1 µSv/h...1 Sv/h

AT3509B Ḣp(10), Ḣp(0.07)

0.1 µSv/h...1 Sv/h

AT3509C Ḣp(10), Ḣp(0.07)

0.1 µSv/h...5 Sv/h

Intrinsic relative error of dose measurement without associated beta radiation

±15% max.

Intrinsic relative error of dose rate measurement

 

0.1 µSv/h...1 µSv/h

±30% max.

1 µSv/h...1 Sv/h

±15% max.

1 Sv/h...5 Sv/h (AT3509C)

±(15+0.001 Ḣp)% max.,

where Ḣp is dose rate in µSv/h

Calibration error for 137Cs

±5%

Energy range

 

AT3509, AT3509B,C

15 keV...10 MeV

AT3509A

30 keV...10 MeV

Energy dependence relative to 662 keV (137Cs)

 

Hp(10) in the following energy range:

 

15 keV...1.5 MeV

±25%

1.5 MeV...10 MeV

±60%

Hp(0.07) in the following energy range (AT3509B,C):

 

15 keV...300 keV

±30%

Alarm thresholds

1 of 8 independent dose thresholds, 1 of 8 independent dose rate thresholds

Anisotropy in angular spacing ±60°

 

For 137Cs and 60Co

±20%

For 241Am

±50%


 



PRICE:
884.95 €
VAT included

On Request

Your message has been sent successfully.
The operation could not be completed.
Enter your first name
Note: All fields are required
Enter your Phone
Enter your message
  • Technical Specifications
  • User Manual
  • Attached Files
  • Related Info
ΠΡΟΤΕΙΝΟΜΕΝΑ ΠΡΟΙΟΝΤΑ